Appropriate chest compression (CC) depth is associated with improved CPR outcome. CCs provided in hospital are often conducted on a compliant mattress.
The objective was to quantify the effect of mattress compression on the assessment of CPR quality in children.
METHODS:
A force and deflection sensor (FDS) was used during CPR in the Pediatric Intensive Care Unit and Emergency Department of a children's hospital.
The sensor was interposed between the chest of the patient and hands of the rescuer and measured CC depth.
Following CPR event, each event was reconstructed with a manikin and an identical mattress/backboard/patient configuration. CCs were performed using FDS on the sternum and a reference accelerometer attached to the spine of the manikin, providing a means to calculate the mattress deflection.
RESULTS:
Twelve CPR events with 14,487 CC (11 patients, median age 14.9 years) were recorded and reconstructed: 9 on ICU beds (9296 CC), 3 on stretchers (5191 CC). Measured mean CC depth during CPR was 47+/-8mm on ICU beds, and 45+/-7 mm on stretcher beds with overestimation of 13+/-4mm and 4+/-1mm, respectively, due to mattress compression.
After adjusting for this, the proportion of CC that met the CPR guidelines decreased from 88.4 to 31.8% on ICU beds (p<0.001), and 86.3 to 64.7% on stretcher (p<0.001). The proportion of appropriate depth CC was significantly smaller on ICU beds (p<0.001).
CONCLUSION:
CC conducted on a non-rigid surface may not be deep enough. FDS may overestimate CC depth by 28% on ICU beds, and 10% on stretcher beds.
2009-05-04
Eng.
Resuscitation
Department of Anesthesiology and Critical Care Medicine, The Children's Hospital of Philadelphia (CHOP), USA. NISHISAKI [at] email.chop.edu
Resuscitation. 2009 May;80(5):540-5
© Galenicom 1999-2013