Assessment of a new Goldmann applanation tonometer.


Abstract

The new GAT digital is as reliable and safe as GAT standard. IOP values correlate well.

It offers a digitised display and a wireless transfer of data.

The display of values up to the first decimal digit is not necessarily associated with a more precise measurement, but may offer an additional comfort compared with the 2 mm Hg scale of the classic GAT.


Full Text


Subjects


Similar articles

Authors


Publication date

2011-12-13


Journal

The British journal of ophthalmology
Br J Ophthalmol (1468-2079)

Journal topics


Language

Eng.


Copyright

The British journal of ophthalmology

Department of Ophthalmology, University of Bern, Inselspital, Bern 3010, Switzerland.


Release reference

Br J Ophthalmol. 2012 Jan;96(1):42-6