Case report of fatal complication in prostatic cryotherapy. First reported death due to argon gas emboli.


Abstract

We present the first reported fatality from argon gas emboli during prostate cryosurgery.

The decedent underwent cryotherapy for prostate carcinoma using cryoablation probes which were cooled with argon and nitrous oxide and warmed with helium.

Minutes into the procedure he experienced sudden cardiovascular collapse and could not be resuscitated.

Postmortem examination was performed at the request of family and healthcare providers.

Collection of tissues and blood samples had to be conducted carefully to capture suspected noble gases,argon, and helium.

Specimens were submitted to Saint Louis University Forensic Toxicology Laboratory for toxicological examination and for evaluation of the composition of the gas retrieved from the vascular system.Gas chromatography mass spectrometric analyses confirmed argon in blood, brain, liver, and gas retrieved from the aorta.

These samples had significant argon compared with room air also sent for comparison.

The manner of death was accident.

To date, there have been no intraoperative surgical fatalities reported from prostatic cryotherapy.

We report such an unfortunate death to raise awareness in the medical community.

We also describe how to collect and handle blood and tissue samples to submit for toxicological analysis in cases of volatile gas emboli.


Full Text


Subjects


Similar articles

Authors


Publication date

2012-03-22


Journal

The American journal of forensic medicine and pathology
Am J Forensic Med Pathol (1533-404X)

Journal topics


Language

Eng.


Copyright

The American journal of forensic medicine and pathology : official publication of the National Association of Medical Examiners

Department of Pathology, Saint Louis University School of Medicine, St. Louis, MO 63134, USA.


Release reference

Am J Forensic Med Pathol. 2012 Mar;33(1):68-72